Diagnostic methods: mass-spectrometry
Mass-spectrometric methods
Huaxuan Wu1, Yinggai Song1, Stephane Ranque2 and Benjamin J. Stielow3
1Department of Dermatology, Peking University First Hospital, Peking University, Beijing, China; 2Parasitology and Mycology, Faculty of Medical and Paramedical Sciences, Aix-Marseille University, Marseille, France; 3Division of Hygiene and Medical Microbiology, Medical University of Innsbruck, Innsbruck, Austria
MALDI-ToF MS
Background
Matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-ToF MS) is an advanced analytical tool widely employed for rapid identification of microorganisms and for biomolecular analysis. The principle of the technique involves ionization of target proteins with a laser and analysis of resulting fragments by measuring their mass-to-charge ratios through time-of-flight (ToF) detection, generating peptide mass fingerprints for database comparison (Singhal et al., 2015; Wickes & Wiederhold, 2018).
Initially conceptualized decades ago, MALDI-ToF MS became widely adopted in clinical microbiology due to its speed, cost-efficiency, and minimal sample preparation requirements (van Veen et al., 2010; Dingle & Butler-Wu, 2013; Robert et al., 2021). The technology produces highly specific protein spectra, akin to barcodes (referred to as Main Spectra Profiles, or MSPs), enabling species-level identification when compared to reference libraries built from strains with proven identity by DNA-sequencing (Robert et al., 2021). The technique operates with either closed databases or, in non-FDA-approved settings, user-expanded libraries that require rigorous validation (Wickes & Wiederhold, 2018). Since different matrices produce distinct spectral patterns, laboratories must strictly follow the same standardized protocols used to create their reference databases. For modification of established procedures, users have several options: adhering to manufacturer specifications, confirming spectral consistency with existing reference data, or developing new databases specifically tailored to the modified protocol (Tsuchida et al., 2020). Recent technical advances have extended its utility to detecting antifungal susceptibility testing (AFST) and identifying microbes directly from clinical samples (Robert et al., 2021).
MALDI-ToF MS offers significant clinical and analytical advantages, particularly in a dramatic reduction of identification time compared to traditional culture-based, physiological and biochemical methods, enabling swift clinical intervention. Fast-growing fungi may be identified within 24 hours of culturing, while slower-growing species require adjusted timelines based on their growth rates. This rapid and accurate identification of pathogens and opportunists facilitates timely targeted antifungal therapy, significantly improving patient management. However, the method is still primarily applied to pure culture isolates, which is a limitation to a further reduction of turnaround time (Clark et al., 2013). Recent applications permit direct testing of clinical samples (REF).
Beyond microbial identification, MALDI-ToF MS can also be applied to detect the culprit in clinical specimens such as blood, cerebrospinal fluid, and urine without prior culture, further enhancing diagnostic efficiency, improving patient prognosis, and reducing hospitalization duration (Patel, 2015; Tsuchida et al., 2020). As a result, it has become an indispensable tool in clinical microbiology. Ongoing advancements, including the expansion of reference databases, refinement of analytical protocols, and integration with complementary technologies like flow cytometry, continue to broaden its clinical applications in mycology.
While MALDI-ToF MS has become a standard tool for yeast identification in many clinical laboratories, its application to filamentous fungi remains constrained by several factors: inadequate reference databases, suboptimal sample preparation protocols, intrinsic variability of colonies, insufficient spectral data, enormous extant diversity, and lack of standardized methods. These limitations often lead to misidentification or failed identification, particularly at the species level. Only matches within the manufacturer’s threshold window are considered to be reliable. Numerous genera of Asco– and Basidiomycota have insufficiently been analyzed and therefore require sequencing with recommended barcoding genes (Sacchi et al., 2002; Tsuchida et al., 2020). Spectral variability due to inconsistent culture conditions (Rizzato et al., 2015) and limited availability of comparative data of fungal pathogens highlight the need for further optimization in database expansion, standardization, and sample processing to fully realize its clinical potential for fungal identification.
Methodology
MALDI-TOF MS analysis involves a carefully orchestrated sequence of steps. The technique begins with sample preparation, where the analyte is mixed with an energy-absorbing organic matrix, such as α-cyano-4-hydroxycinnamic acid (CHCA), 2,5-dihydroxybenzoic acid (DHB), sinapinic acid (SA), or ferulic acid (FA) (Fenselau & Demirev, 2001; Vaidyanathan et al., 2002; Williams et al., 2003a; Croxatto et al., 2012). Upon drying, the sample co-crystallizes within the matrix, forming a homogeneous mixture suitable for analysis. The matrix plays a critical role by absorbing laser energy, facilitating sample desorption and ionization while preventing analyte fragmentation. When irradiated with a pulsed nitrogen laser, the matrix-sample mixture undergoes desorption, generating predominantly singly protonated ions (Yates, 1998). These ions are then accelerated through an electrostatic field and travel through a vacuum flight tube, where their time-of-flight (ToF) is measured and correlated with their mass-to-charge ratio (m/z). Advanced instruments may incorporate an ion mirror to enhance resolution by correcting for kinetic energy variations among ions (Yates, 1998). The resulting mass spectrum, or peptide mass fingerprint (PMF), serves as a unique molecular signature for identification by comparison against reference databases.
MALDI-ToF MS instrumentation comprises three main components: an ion source, a mass analyzer, and a detection system. The ion source utilizes soft ionization techniques like MALDI, which are particularly effective for preserving large biomolecules such as fungal proteins during ionization (Emonet et al., 2010). Time-of-flight (ToF) analyzers, the most commonly used mass analyzers in fungal and microbial studies, measure ion flight times through a vacuum tube (with smaller ions arriving faster) and are favored for their rapid analysis capabilities and compatibility with pulsed laser ionization (Heller et al., 1987). To enhance analytical performance, advanced hybrid configurations, including tandem (ToF-ToF), hybrid quadrupole-ToF (Q-ToF), and multistage systems like Q-Q-ToF and triple quadrupole, combine the strengths of different analyzers to improve resolution and accuracy for complex samples (Domon & Aebersold, 2006). The detection system records ion m/z values and intensities, generating spectra with distinct peaks typically ranging from 1,000 to 20,000 m/z (Croxatto et al., 2012). The choice of mass analyzer hinges on both sample characteristics (e.g., complexity, molecular composition such as proteins or lipids) and analytical objectives (e.g., quantification, microbial identification, or structural profiling) (Croxatto et al., 2012; Tsuchida et al., 2020).
Sample preparation for fungal analysis may require additional pretreatment steps to ensure efficient lysis and protein extraction. While vegetative fungal cells are generally lysed by the organic solvents and acids present in the MALDI matrix, more resistant forms such as yeast cells may necessitate pretreatment with strong organic acids or alcohols to facilitate adequate protein release (Bizzini et al., 2010). Standardized protocols are essential to maintain consistency between experimental results and database entries, as even minor deviations in sample preparation or matrix selection can alter spectral profiles and compromise identification accuracy.
For fungal identification, MALDI-ToF MS relies on comparing the obtained PMF against extensive reference databases or matching specific biomarker masses to predicted proteomic data (Singhal et al., 2015). The mass range between 2−20 kDa is particularly significant as it encompasses ribosomal proteins and other housekeeping proteins that constitute approximately 60−70% of the fungal cellular dry mass (Murray, 2012). The highly conserved nature of these proteins allows for reliable genus- or species-level classification when their spectral patterns are aligned with database entries (Fagerquist et al., 2010). The selection of an appropriate matrix is crucial for optimal results, with CHCA and DHB being particularly effective for detecting lower mass ions below 10 kDa (Hathout et al., 2000; Williams et al., 2003b; Ruelle et al., 2004; Giebel et al., 2010), while SA and FA are better suited for higher mass proteins exceeding 15 kDa, albeit with reduced sensitivity (Madonna et al., 2000; Vargha et al., 2006).
Application
MALDI-ToF MS has become an indispensable tool in clinical microbiology for the rapid and accurate identification of microorganisms, including fungi. While its application for identification of bacteria and yeasts is well-established, its use for filamentous fungi is still under development, supported by continuous improvements in instrumentation, database expansion, and sample preparation protocols. Several commercial MALDI-ToF MS systems are currently employed in clinical and research settings, of which the Bruker Biotyper system, the bioMérieux VITEK MS series, and the Autobio Autof MS systems are the most commonly used. Additionally, non-commercial databases such as MicrobeNet, MSI and some in-house databases have been developed to enhance the identification of rare and cryptic species. This section evaluates these platforms, focusing on their performance, operational workflows, specimen compatibility, and comparative advantages and limitations.
The Bruker Biotyper system is widely utilized for fungal identification, supporting both yeast and mold analyses. It generates identification scores ranging from 0 to 3, with values below 1.7 considered unreliable, scores between 1.7 and 2.0 indicating genus-level identification, and scores above 2.0 providing species-level resolution. The system employs a reusable target plate and a formic acid-acetonitrile extraction protocol for mold processing. Performance of the system is heavily dependent on its database. Evaluations have shown that the Bruker system achieves high genus-level identification rates for common fungi, though species-level accuracy can be challenging for cryptic species within complexes such as the Aspergillus sections Fumigati or Nidulantes (Dutkiewicz et al., 2025). A significant limitation noted in some database versions is the absence of spectra for thermally dimorphic fungi like Histoplasma and Coccidioides, which restricts its utility in regions where these pathogens are endemic (Ransom et al., 2025).
A large-scale national surveillance study in China evaluating 2683 yeast isolates demonstrated that the Bruker Biotyper MS system achieved a high overall identification accuracy of 98.8% (compared to 95.4% for the Vitek MS system; Wang et al., 2016). The performance was particularly successful with Candida and candida-like species, reaching an accuracy of 99.4% (Wang et al., 2016). In another study, the Bruker Biotyper achieved 98.7% identification sensitivity for non-albicans Candida and uncommon yeasts (Teke et al., 2021). For filamentous fungi, performance is more variable. For example, Sun et al. (2021) demonstrated 42.1% species-level identification for clinical moulds and Tang et al. (2025) 72% for Nannizzia isolates when using manufacturer-recommended protocols.
The system’s flexibility is a key feature. It allows for manual reflexing of spectra acquisition when automated methods fail, which can improve identification success (Ransom et al., 2025). However, its workflow may involve longer hands-on time compared to some newer systems, particularly when formic acid treatment is applied universally to all isolates (Bosserman et al., 2025). Studies comparing the Bruker Biotyper with other systems indicate that it demonstrates a very low rate of misidentification, which is critical for clinical safety (Hankins et al., 2023; Lévesque et al., 2015).
Sample preparation ia another determinant for successful identification. The system supports three main sample preparation methods: direct, on-plate, and in-tube extractions, with average identification rates of 83.8%, 96.0%, and 95.3%, respectively, for clinical isolates (Wang et al., 2021). The in-tube method generally yields the highest sensitivity, although no statistically significant difference was observed between the methods (Choi et al., 2022). The choice of method may impacts results, as exemplified by an improved identification accuracy for Cryptococcus species using the recommended protein extraction method as opposed to direct on-plate method (Wang et al., 2016). For dermatophyte identification, the Bruker system’s performance can be significantly improved by using selective culture media such as ID fungi plate (IDFP) medium, and its accuracy is notably enhanced when supplemented with the MSI database, which outperforms the Bruker library for this group of fungi (Jeraldine et al., 2023). The use of IDFP medium has been shown to significantly increase identification rates for filamentous fungi with the Bruker system, with improvements of 22% for genus-level and 8% for species-level identification compared to standard Sabouraud’s glucose agar (Heireman et al., 2020).
The VITEK MS platform, particularly the newer PRIME model, incorporates workflow enhancements such as a continuous load-and-go mechanism, an urgent CHECK slide analysis option, reducing slide processing time (Bosserman et al., 2025). The VITEK MS system presents results as percentage probabilities (60−99%), where higher values denote greater confidence, and scores below 60% are deemed inconclusive. VITEK MS PRIME delivers the same identification performance as VITEK MS, yet reduces hands-on time by 60−66% and process completion time by 50−51%, owing to its higher laser frequency, faster spectrum acquisition and continuous ‘load-and-go’ sample loading that eliminates slide tray handling (Bardelli et al., 2022). The system uses disposable target slides and offers two primary spotting options: a plastic loop or the proprietary PICKME pen. For fungal identification, bioMérieux recommends a direct transfer technique for yeasts and a standardized chemical extraction using the VITEK MS Mold Kit for filamentous fungi (Dutkiewicz et al., 2025).
In performance evaluations, the VITEK MS system demonstrated 89.2% correct identification of closely-related yeast species, with particular limitations observed in identifying less common species within species complexes (Yi et al., 2021). However, another extensive evaluation reported a correct species-level identification rate of 93.7% for the VITEK MS system, no statistically significant difference being observed with the Bruker platform (Deak et al., 2015). For filamentous fungi, a multicenter evaluation using the VITEK MS v3.2 system with the manufacturer-recommended extraction protocol, the platform achieved a high species-level identification rate of 96.0% for clinically relevant moulds (Sun et al., 2021). In contrast, a study utilizing the earlier VITEK MS v3.0 system reported that 82.2% of isolates were identified to the genus/complex/species group level, with the species-level resolution within that group being 63.5% (Pinheiro et al., 2019). This disparity reflects the performance of the updated database. The VITEK MS PRIME has demonstrated comparable performance to the Bruker system in identifying common yeasts and molds, with species-level identification rates exceeding 90% for yeasts using the IVD database (v3.2)(Lee et al., 2024). However, the system’s acquisition software includes a stringent quality control algorithm that can result in a higher proportion of low-quality outputs, potentially necessitating repeated extraction or instrument tuning (Dutkiewicz et al., 2025). In side-by-side comparisons, the VITEK MS PRIME showed advantages in reducing hands-on time and achieving a faster time-to-result than the Bruker system, especially in high-throughput laboratory environments (Bosserman et al., 2025). User experience assessments have indicated a preference for the VITEK MS system in some laboratory settings, e.g. ease of use and workflow integration (Deak et al., 2015).
While the VITEK commercial database coverage for frequently encountered pathogens is robust, it, like other systems, demonstrates lower performance for rare species poorly represented in the library. The use of the SARAMIS RUO database (v4.16) significantly improves identification, particularly for basidiomycetes, raising species-level identification from 35.3% to 100% in one study (Lee et al., 2024). For filamentous fungi, the VITEK MS system has shown a high correct identification rate for isolates within its database but a concerning rate of misidentification for those not included, which necessitates careful verification of results for uncommon species (Hankins et al., 2023). This issue was particularly evident in a study where VITEK MS generated incorrect identifications for 38.7% of strains not present in its database (Lévesque et al., 2015).
The Autobio Autof MS systems, particularly the MS1000 and MS2600 models, are increasingly utilized in clinical microbiology, with significant adoption in China. For fungal identification, the system employs a standardized scoring threshold where a log score ≥9.0 indicates secure species-level identification and a score between 6.0 and 8.99 indicates genus-level identification. In a real-time clinical evaluation, the Autof MS2600 demonstrated exceptional performance by correctly identifying 98.02% of yeast isolates at a log score ≥9.0 (rising to 99.01% at ≥7.0) without any database expansion or threshold adjustments, significantly outperforming the MBT smart system which required both an in-house updated database and lowered score thresholds to achieve comparable results (De Carolis et al., 2024). Further validating its utility in direct specimen testing, a prospective study of 125 positive blood cultures reported that the Autof MS1000 accurately identified all 12 yeast isolates, including Candida auris, C. tropicalis, and C. parapsilosis, achieving 100% concordance with conventional methods (Benwan et al., 2025). The system has also exhibited high performance in differentiating closely related yeast species, achieving 99.2% correct identification compared to 89.2% for the VITEK MS system (Yi et al., 2021).
For the identification of filamentous fungi, the Autof MS1000 system demonstrates high performance, although its efficacy is significantly influenced by the sample preparation method. One study utilizing an FA-sandwich pretreatment reported genus- and species-level identification rates of 95.9% and 93.9%, respectively (Peng et al., 2022). This method, which requires only a 1 mm diameter sample and reduces processing time to 5-10 minutes per isolate, provides a practical and efficient alternative to more complex full extraction protocols. The system showed robust performance for common clinical moulds, achieving species-level identification rates of 97.1% for Aspergillus, 100% for Fusarium, and 100% for Trichophyton, and successfully identified rare species like Talaromyces marneffei when reference spectra were available (Peng et al., 2022).
An evaluation using a simplified formic acid extraction method reported high genus-level (95.28%) but lower species-level (76.41%) identification rates (Ao et al., 2023). However, database coverage was noted as a limitation, preventing the reliable identification of rare species in Beauveria, Cunninghamella, and Fonsecaea (Ao et al., 2023). The critical impact of the extraction method is further underscored by reports that tube extraction (59% identification rate) outperforms on-plate extraction (51%) for filamentous fungi (Sun et al., 2021). When optimal preparation methods are employed, such as the on-target extraction method, the platform achieved up to 100% correct identification, as shown in an evaluation of 321 clinical isolates (Wang et al., 2021).
Operationally, the Autof software is noted for its user-friendly interface, integrating multiple functions into a single platform, which simplifies operation compared to systems requiring multiple interfaces (Ao et al., 2023). In a comparison with the EXS3000 system, both instruments showed similar overall identification performance for microorganisms including molds, though the EXS3000 had a slight advantage in species-level mold identification (81.82% vs. 78.79%) and used a reusable target plate, reducing long-term costs (Xiong et al., 2023). When evaluated against the Bruker Biotyper, the Autof MS1000 demonstrated comparable performance, with species-level identification rates of 98.5% versus 97.6% for preserved strains and 99.7% versus 99.3% for clinical isolates (Park et al., 2021). Additionally, the Autof system processed 96 samples in approximately 14 minutes, significantly faster than the 27 minutes required by the Bruker system (Park et al., 2021).
The ASTA MicroIDSys MALDI-TOF platform, developed by Korea-based ASTA Inc., couples a solid-state laser and modular hardware with dedicated databases such as MicroIDSys CoreDB and MycoDB. Using a direct-smear protocol and an acceptance threshold of ≥140, it delivers rapid microbial identification and has proven effective for yeasts and other clinically relevant fungi. For yeast identification, it demonstrated 95.2% concordance with Bruker Biotyper for Candida species at the species level (Chung et al., 2021). The system shows comparable performance to established platforms and may offer a cost-effective alternative for resource-limited settings.
Non-commercial and in-house databases
Beyond commercial systems, public and custom databases are vital for comprehensive fungal identification. The MSI database is an open-access, non-commercial resource that has proven highly valuable for identifying rare and cryptic fungi, particularly for Aspergillus species and rare molds, where it frequently outperforms commercial libraries (Dutkiewicz et al., 2025). The platform hosts an extensive database maintained by the BCCM/IHEM culture collection and Marseille Hospital Mycology Laboratory, encompassing 1702 species across multiple genera and enables significantly faster and more accurate species-level identification of diverse fungi compared to commercial systems (Normand et al., 2017).
The newly developed MSI-2 application demonstrated superior performance compared to both the previous MSI-1 version and the Bruker filamentous fungi database (Normand et al., 2021). Using a challenging panel of 633 mould isolates previously identified by DNA-based methods, MSI-2 achieved 83.25% correct species-level identification, significantly outperforming MSI-1 (63.19%) and the Bruker system for the same species with a 1.7 threshold (38.07%)(Normand et al., 2021). The improvement was particularly notable for Aspergillus and Fusarium, including many cryptic species, with MSI-2 reaching 90% and 78% correct identifications for these genera, respectively (Normand et al., 2021). A key advantage is its compatibility with both Bruker and bioMérieux instruments, providing flexibility for laboratories regardless of their platform (Dutkiewicz et al., 2025). Its utility, however, is dependent on the use of proper protein extraction protocols, as the database was constructed using spectra from fully extracted samples and may not be compatible with spectra generated from direct deposition methods. For dermatophyte identification, the MSI database has demonstrated superior performance compared to the Bruker MBT library, achieving higher correct identification rates across different culture media (Jeraldine et al., 2023). In one evaluation, the MSI library provided coverage for 92% of tested fungal isolates compared to only 73% for the Bruker library, significantly enhancing identification capabilities for both dermatophytes and non-dermatophyte filamentous fungi (Heireman et al., 2020).
An alternative to public databases is the creation of in-house libraries. To overcome the limitations of commercial MALDI-ToF MS databases for fungal identification, laboratories can construct their own in-house reference databases using well-characterized clinical isolates, or utilize publicly accessible online platforms. The process of building a reliable in-house has the following step. Initially, every spectrum undergoes quality screening: a minimum of twenty-five peaks must exhibit resolution above four hundred, and at least two of those peaks must surpass five hundred (Becker et al., 2014a). After validation, multiple spectra are merged into a single representative metaspectrum (MSP) through the instrument’s own software. This consolidation typically draws on four distinct biological cultures, each analyzed in ten technical replicates, yielding a robust consensus profile (Cassagne et al., 2011).
Several online platforms facilitate identification. One such resource is MicrobeNet (https://microbenet.cdc.gov), which offers a downloadable list of bacterial and fungal species included in its MALDI database and specifies the origin of each entry.
In addition to these web platforms, numerous laboratories have developed specialized in-house databases to improve identification coverage. Notable examples include the NIH mould database (Lau et al., 2013), along with those developed by Gautier et al. (2014) and Becker et al. (2014b), which cover 152, 347, and 472 fungal species with reported identification accuracies of 87.9%, 98.8%, and 95.5%, respectively (Wilkendorf et al., 2020). Furthermore, a database developed by Oliver Bader’s laboratory is freely available and can be integrated into local systems: it has been validated for identifying species in Scedosporium (Bernhard et al., 2016) and the Cryptococcus neoformans (Bernhard et al., 2021). Expanding commercial databases with in-house data can substantially improve correct identification rates, particularly for species frequently encountered in clinical practice.
Reproducibility and workflow considerations
The accuracy of any database, commercial or custom, is paramount, but the accuracy of in-house database entries is critical for reliability. Omission of type and reference strains can lead to erroneous identification (Wilkendorf et al., 2020). This underscores the importance of using correctly annotated reference strains, as errors in database entries can propagate. For example, Meyerozyma caribbica previously was misidentified due to an erroneously labeled reference spectrum in the CBS database attributed to M. guilliermondii (Quintilla et al., 2018). Once reference entries were corrected, distinction of the two species proved feasible. MALDI-ToF also misidentifies Candida sake DSM 70763 as C. oleophila because its reference spectrum in the commercial BDAL v.4.0 database is incorrectly labeled, as revealed by sequence re-validation (Quintilla et al., 2018). A study by Wilkendorf et al. (2020) evaluating the performance of in-house MALDI-TOF MS databases for the identification of filamentous fungi reported identification rates ranging from 82.1% to 100% for the species studied. This indicated that modest augmentation of commercial libraries has been shown to elevate identification accuracy substantially, particularly when a laboratory encounters the same species repeatedly. Consequently, every strain incorporated into an institutional reference collection must be authenticated with the utmost rigor; any misclassification of a repository isolate will propagate directly to subsequent clinical specimens matched against that record. Due to rapid changes in fungal taxonomy, in-house databases should be regularly reviewed and updated as necessary to ensure accurate identification according to recent literature. Ongoing research continues to yield new in-house databases for MALDI-TOF MS-based fungal identification. However, most of these databases are not yet publicly available. Authors typically list the strains used for database construction but seldom provide the corresponding MALDI spectra or profiles necessary for direct comparison. As summarized in Table 1, a list of selected yeast-related databases is available for further reference.
Comparative Performance and Workflow Considerations
The two most widely adopted MALDI-ToF MS platforms for microbial identification, the MALDI Biotyper (Bruker Daltonics, Bremen, Germany) and the VITEK MS (bioMérieux, Marcy-l’Étoile, France), employ similar analytical principles but differ significantly in database composition, algorithmic processing, and result interpretation (WHO, 2025). The MALDI Biotyper system generates identification scores ranging from 0 to 3, with values below 1.7 considered unreliable, scores between 1.7 and 2.0 indicating genus-level identification, and scores above 2.0 providing species-level resolution. In contrast, the VITEK MS system presents results as percentage probabilities (60−99%), where higher values denote greater confidence, and scores below 60% are deemed inconclusive. These methodological distinctions influence their performance in identification of clinical fungi.
Comparative studies have demonstrated that the MALDI Biotyper exhibits superior accuracy in identifying ascomycetous yeasts, achieving correct identification rates of 98.8% compared to 95.4% for VITEK MS (Wang et al., 2016). This advantage extends to Candida and candida-like species, where the Bruker system attains 99.4% accuracy versus 95.5% for VITEK MS (Wang et al., 2016). The discrepancy is partly attributed to the broader taxonomic coverage of the Bruker database, which includes a more extensive collection of fungal species, particularly filamentous fungi. Additionally, while VITEK MS offers an alternative Saramis database with expanded species coverage, its performance remains understudied compared to the standard IVD database (Lévesque et al., 2015).
Methodological differences further distinguish the systems. The Bruker system liquid cultures grown for maximally 2 days and necessitates a formic acid extraction step (either on-slide or in-tube). The VITEK MS system offers a more streamlined workflow with direct on-plate formic acid treatment, demonstrating greater tolerance to sample variability and generating fewer unusable spectra, though with slightly lower overall identification accuracy (Deak et al., 2015). These differences suggest the Bruker system may be preferable when maximum accuracy is critical, while VITEK MS provides greater efficiency and convenience for routine clinical use.
When comparing the workflow of both systems, the VITEK system offers advantages, such as faster target setup and lower overall hands-on time (Bosserman et al., 2025). A detailed workflow analysis found that Bruker’s workflow shortens hands-on time by 3-6 min and the Sirius measurement reduces TTR three- to five-fold, whereas VITEK MS PRIME cuts user interaction to ≈1.5 min per target. With no significant difference in TAT during target preparation between the two systems, the former is ideal for large-batch rapid batching and the latter for intermittent parallel loading (Thelen et al., 2023). The Autof ms1000 excels in analysis speed and operational simplicity but requires further database expansion for comprehensive mold identification (Ao et al., 2023; Xiong et al., 2023). The ASTA MicroIDSys system shows comparable performance to established platforms for routine isolates but may have limitations with rare species (Chung et al., 2021).
Detection limit
While identification of yeasts by MALDI-ToF has become routine in many clinical laboratories, the analysis of filamentous fungi often presents greater challenges and requires more specialized protocols. Even for yeasts, however, reliable identification is not always guaranteed. For instance, while the technique reliably assigns isolates to the Kazachstania telluris species complex, it fails to resolve individual species within this group, such as K. bovina and its closely related siblings, necessitating additional ITS sequencing for precise differentiation (Kaeuffer et al., 2022). Similarly, the cryptic species Candida africana is occasionally misidentified as C. albicans, underscoring the difficulty in distinguishing these two species (Jonani et al., 2024). In the case of Candida magnoliae, MALDI-ToF peptide mass fingerprinting was severely limited by low sequence coverage and poor cross-species protein homology, forcing reliance on MS/MS and de novo sequencing for reliable identification (Kim et al., 2004).
The choice of sample preparation method also significantly impacts results, with complete protein extraction consistently yielding more reliable identifications for molds than direct deposition (Dutkiewicz et al., 2025; Ransom et al., 2025). In a direct comparison of three MALDI-TOF MS systems for filamentous fungal identification, the Bruker Biotyper showed the highest correct identification rate (71.43%), followed by ASTA MicroIDSys (70.24%) and VITEK MS (55.95%) when using the in-tube extraction method (Choi et al., 2022). The study emphasized that differences in sensitivity were primarily attributable to variations in database composition across the systems rather than intrinsic instrument capabilities. For filamentous fungi, the use of selective culture media such as IDFP or SGC2 agar has been shown to significantly improve identification rates compared to non-selective media like Sabouraud’s glucose agar, regardless of the MALDI-TOF MS system used (Jeraldine et al., 2023). Furthermore, the optimal incubation period for fungal growth before analysis varies by group, with dermatophytes generally requiring longer incubation (3-4 days) than Aspergillus (2-3 days) to achieve optimal identification rates (Jeraldine et al., 2023). The implementation of rapid protocols such as the Rapid Sepsityper® can significantly reduce turnaround time for yeast identification from positive blood cultures, achieving reliable identification in as little as 10 minutes with the MBT-Sepsityper module (Ponderand et al., 2020).
Comparison with Conventional and Molecular Methods
MALDI-ToF MS offers substantial advantages over conventional phenotypic identification, which often requires extended incubation for colony maturation and may only achieve genus-level resolution (Ransom et al., 2025). In contrast, MALDI-ToF MS can provide species-level identification within minutes to hours, facilitating more timely clinical decision-making (Bosserman et al., 2025). The technology’s effectiveness is further enhanced when supplemented with public databases like MSI or MicrobeNet, which provide expanded coverage for uncommon species that may be absent from commercial libraries. The integration of MALDI-ToF MS into a combined diagnostic protocol can dramatically reduce reporting times. One study demonstrated that a protocol combining MALDI-ToF MS for rapid identification directly from positive culture broths with subsequent rapid antimicrobial susceptibility testing (RAST) reduced the turnaround time by 18 to 36 hours compared to standard methods (Tian et al., 2016). This approach facilitates earlier optimization of antimicrobial therapy, which is crucial for patient management. The technology enables rapid differentiation of clinically relevant species with distinct antifungal susceptibility profiles, directly impacting treatment selection. However, its performance remains intrinsically linked to the quality and comprehensiveness of its spectral databases. Molecular methods, such as DNA sequencing of the ITS region or other barcodes, remain the reference standard for resolving discrepant identifications or characterizing rare species not covered by MALDI-ToF MS databases. MALDI-ToF MS systems demonstrated significantly higher accuracy compared to phenotypic identification methods, with an average of only 2.76% misidentification or no-identification rates for MS systems versus 8.50% for phenotypic methods (Bilecen et al., 2015).
Practical identification
The Bruker MALDI Biotyper and bioMérieux VITEK MS systems both maintain extensive databases of clinically relevant yeasts and filamentous fungi, but there are discrepancies in nomenclature. For example, Bruker lists the sexual name Cyberlindnera jadinii, while VITEK MS uses its asexual name Candida utilis. Although the optimal naming of these fungi is still under scrutiny, but the connection of both and the currently recommended name can be found at www.atlasclinicalfungi.org/nomenclaturenames. The MALDI Biotyper CA System, developed by Bruker Daltonics, represents a clinical IVD-compliant version of the MALDI-ToF platform specifically designed to meet U.S. regulatory requirements. Both systems provide FDA-approved databases containing clinically relevant yeast species, though each includes some unique entries (Carroll & Patel, 2018; Patel, 2019).
Clinically relevant yeasts in references / databases dd. August 2025.
Asco- and basidiomycetous yeasts in the literature. The used database in each reference is listed in the first column.